T1 - Synchronized Desorption Electrospray Ionization Mass Spectrometry Imaging
AU - Comi, Troy J.
AU - Ryu, Seung Woo
AU - Perry, Richard H.
N1 - The Perry Research Laboratory acknowledges financial support from the University of Illinois at Urbana-Champaign (UIUC), NIH Training Program at Chemistry-Interface with Biology (NIH T32 GM070421), National Science Foundation Graduate Research Fellowship Program, and the UIUC Springborn Fellowship. We also thank Dr. Edward T. Chainani for help in designing initial circuitry, assisting with valve latency measurements, and many useful conversations.
PY - 2016/1/19
Y1 - 2016/1/19
N2 - Desorption electrospray ionization (DESI) has emerged as a powerful technique for mass spectral analysis and imaging under ambient conditions. Synchronization of DESI (sDESI) with the ion injection period (IT)of low-duty cycle mass spectrometers has been previously shown to improve sensitivity and reduce the amount of sample depleted during the acquisition of each spectrum (viz. MS scan time). In this report, we describe the development and characterization of an sDESI mass spectrometry imaging source (sDESI-MSI). Our results show that synchronization of DESI with the IT of an LTQ Orbitrap-XL mass spectrometer improves spatial resolution by factors of ∼4-6. In addition, under certain experimental conditions, synchronization was essential to acquire distinct MS images of low-intensity endogenous FAs (< 5% relative intensity) in fingermarks at high sampling frequencies (step sizes ≤ 75 μm). The magnitudes of these improvements in performance depend on the properties of the microdroplet spray, sample, and surface. Simulations that model analyte movement during desorption and the "washing effect" replicate the experimental results with the washing parameter having the greatest impact on performance. Thus, synchronization improves spatial resolution and sensitivity by decreasing the percentage of the total MS scan time that analytes are influenced by the "washing effect". Generally, synchronization of DESI with IT improves performance and expands the range of analytes, surfaces, and experimental conditions amenable to DESI-MSI, especially for analytes that are weakly attached to a surface.
AB - Desorption electrospray ionization (DESI) has emerged as a powerful technique for mass spectral analysis and imaging under ambient conditions. Synchronization of DESI (sDESI) with the ion injection period (IT)of low-duty cycle mass spectrometers has been previously shown to improve sensitivity and reduce the amount of sample depleted during the acquisition of each spectrum (viz. MS scan time). In this report, we describe the development and characterization of an sDESI mass spectrometry imaging source (sDESI-MSI). Our results show that synchronization of DESI with the IT of an LTQ Orbitrap-XL mass spectrometer improves spatial resolution by factors of ∼4-6. In addition, under certain experimental conditions, synchronization was essential to acquire distinct MS images of low-intensity endogenous FAs (< 5% relative intensity) in fingermarks at high sampling frequencies (step sizes ≤ 75 μm). The magnitudes of these improvements in performance depend on the properties of the microdroplet spray, sample, and surface. Simulations that model analyte movement during desorption and the "washing effect" replicate the experimental results with the washing parameter having the greatest impact on performance. Thus, synchronization improves spatial resolution and sensitivity by decreasing the percentage of the total MS scan time that analytes are influenced by the "washing effect". Generally, synchronization of DESI with IT improves performance and expands the range of analytes, surfaces, and experimental conditions amenable to DESI-MSI, especially for analytes that are weakly attached to a surface.
UR - http://www.scopus.com/inward/record.url?scp=84955310644&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84955310644&partnerID=8YFLogxK
U2 - 10.1021/acs.analchem.5b03010
DO - 10.1021/acs.analchem.5b03010
M3 - Article
C2 - 26569449
AN - SCOPUS:84955310644
SN - 0003-2700
VL - 88
SP - 1169
EP - 1175
JO - Analytical Chemistry
JF - Analytical Chemistry
IS - 2
ER -