T1 - Repeatability and accuracy of fingertip pulse oximeters for measurement of hemoglobin oxygen saturation in arterial blood and pulse rate in anesthetized dogs breathing 100% oxygen
AU - Ambrisko, Tamas D.
AU - Dantino, Stephanie C.
AU - Keating, Stephanie C.J.
AU - Strahl-Heldreth, Danielle E.
AU - Sage, Adrianna M.
AU - Martins, Felipe Da Costa
AU - Harper, Tisha A.M.
AU - Wilkins, Pamela A.
N1 - Publisher Copyright:
© 2021, American Veterinary Medical Association. All rights reserved.
PY - 2021/4
Y1 - 2021/4
N2 - OBJECTIVE To evaluate the repeatability and accuracy of fingertip pulse oximeters (FPO) for measurement of hemoglobin oxygen saturation in arterial blood and pulse rate (PR) in anesthetized dogs breathing 100% O2. ANIMALS 29 healthy client-owned anesthetized dogs undergoing various surgical procedures. PROCEDURES In randomized order, each of 7 FPOs or a reference pulse oximeter (PO) was applied to the tongue of each intubated anesthetized dog breathing 100% O2. Duplicate measurements of oxygen saturation (Spo2) and PR were obtained within 60 seconds of applying an FPO or PO. A nonparamet-ric version of Bland-Altman analysis was used. Coefficient of repeatability was the interval between the 5th and 95th percentiles of the differences between duplicate measurements. Bias was the median difference, and the limits of agreement were the 5th and 95th percentiles of the differences between each FPO and the PO. Acceptable values for the coefficient of repeatability of Spo2 were ≤ 6%. Agreements were accepted if the limits of agreement had an absolute difference of ≤ ± 3% in Spo2 and relative difference of ≤ ± 10% in PR. RESULTS Coefficient of repeatability for Spo2 was acceptable for 5 FPOs, but the limits of agreement for Spo2 were unacceptable for all FPOs. The limits of agreement for PR were acceptable for 2 FPOs. CONCLUSIONS AND CLINICAL RELEVANCE Results suggested that some FPOs may be suitable for accurately monitoring PRs of healthy anesthetized dogs breathing 100% O2, but mild underes-timation of Spo2 was common. (Am J Vet Res 2021;82:268–273).
AB - OBJECTIVE To evaluate the repeatability and accuracy of fingertip pulse oximeters (FPO) for measurement of hemoglobin oxygen saturation in arterial blood and pulse rate (PR) in anesthetized dogs breathing 100% O2. ANIMALS 29 healthy client-owned anesthetized dogs undergoing various surgical procedures. PROCEDURES In randomized order, each of 7 FPOs or a reference pulse oximeter (PO) was applied to the tongue of each intubated anesthetized dog breathing 100% O2. Duplicate measurements of oxygen saturation (Spo2) and PR were obtained within 60 seconds of applying an FPO or PO. A nonparamet-ric version of Bland-Altman analysis was used. Coefficient of repeatability was the interval between the 5th and 95th percentiles of the differences between duplicate measurements. Bias was the median difference, and the limits of agreement were the 5th and 95th percentiles of the differences between each FPO and the PO. Acceptable values for the coefficient of repeatability of Spo2 were ≤ 6%. Agreements were accepted if the limits of agreement had an absolute difference of ≤ ± 3% in Spo2 and relative difference of ≤ ± 10% in PR. RESULTS Coefficient of repeatability for Spo2 was acceptable for 5 FPOs, but the limits of agreement for Spo2 were unacceptable for all FPOs. The limits of agreement for PR were acceptable for 2 FPOs. CONCLUSIONS AND CLINICAL RELEVANCE Results suggested that some FPOs may be suitable for accurately monitoring PRs of healthy anesthetized dogs breathing 100% O2, but mild underes-timation of Spo2 was common. (Am J Vet Res 2021;82:268–273).
UR - http://www.scopus.com/inward/record.url?scp=85103502295&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85103502295&partnerID=8YFLogxK
U2 - 10.2460/ajvr.82.4.268
DO - 10.2460/ajvr.82.4.268
M3 - Article
C2 - 33764836
AN - SCOPUS:85103502295
SN - 0002-9645
VL - 82
SP - 268
EP - 273
JO - American journal of veterinary research
JF - American journal of veterinary research
IS - 4
ER -