Abstract
An efficiency optimized single photon tomographic system, GAMMATOM-1, is developed and its performance is studied. The system consists of 36 Nal(Tl) crystals of 2cm (diameter) x 3.8cm (length) arranged in a circular form with four banks or arcs (each arc with 9 Nal(Tl) crystals). Resolution of 7mm fwhm is obtained with a 16cm diameter resolution-phantom of 99mtc. Sensitivity of the system has also been measured and obtained 10K ∼ 15K (counts/sec) / (μCi/cc) with a 20cm diameter uniform circular phantom of 99mtc. System design details and performance are reported.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 484-487 |
| Number of pages | 4 |
| Journal | IEEE Transactions on Nuclear Science |
| Volume | 29 |
| Issue number | 1 |
| DOIs | |
| State | Published - Feb 1982 |
| Externally published | Yes |
ASJC Scopus subject areas
- Nuclear and High Energy Physics
- Nuclear Energy and Engineering
- Electrical and Electronic Engineering
Fingerprint