Abstract
We developed a radiation detector based on an organic scintillator for spectrometry and dosimetry of out-of-field secondary neutrons from clinical proton beams. The detector consists of an EJ-299-34 crystalline organic scintillator, coupled by fiber optic cable to a silicon photomultiplier (SiPM). Proof of concept measurements were taken with 137Cs and 252Cf, and corresponding simulations were performed in MCNPX-PoliMi. Despite its small size, the detector is able to discriminate between neutron and gamma-rays via pulse shape discrimination. We simulated the response function of the detector to monoenergetic neutrons in the 100 keV-0 MeV range using MCNPX-PoliMi. The measured unfolded 252Cf neutron spectrum is in good agreement with the theoretical Watt fission spectrum. We determined the ambient dose equivalent by folding the spectrum with the fluence-to-ambient dose conversion coefficient, with a 1.4% deviation from theory. Some preliminary proton beam experiments were preformed at the Bronowice Cyclotron Center patient treatment facility using a clinically relevant proton pencil beam for brain tumor and craino-spinal treatment directed at a child phantom.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 355-359 |
| Number of pages | 5 |
| Journal | Radiation Protection Dosimetry |
| Volume | 180 |
| Issue number | 1-4 |
| DOIs | |
| State | Published - 2018 |
| Externally published | Yes |
ASJC Scopus subject areas
- Radiological and Ultrasound Technology
- Radiation
- Radiology Nuclear Medicine and imaging
- Public Health, Environmental and Occupational Health
Fingerprint
Dive into the research topics of 'Organic scintillator for real-time neutron dosimetry'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS