TY - JOUR
T1 - Non-invasive Deep-Brain Imaging with 3D Integrated Photoacoustic Tomography and Ultrasound Localization Microscopy (3D-PAULM)
AU - Tang, Yuqi
AU - Wang, Nanchao
AU - Dong, Zhijie
AU - Lowerison, Matthew
AU - Del Aguila, Angela
AU - Johnston, Natalie
AU - Vu, Tri
AU - Ma, Chenshuo
AU - Xu, Yirui
AU - Yang, Wei
AU - Song, Pengfei
AU - Yao, Junjie
N1 - Publisher Copyright:
© 1982-2012 IEEE.
PY - 2024
Y1 - 2024
N2 - Photoacoustic computed tomography (PACT) is a proven technology for imaging hemodynamics in deep brain of small animal models. PACT is inherently compatible with ultrasound (US) imaging, providing complementary contrast mechanisms. While PACT can quantify the brain's oxygen saturation of hemoglobin (sO2), US imaging can probe the blood flow based on the Doppler effect. Further, by tracking gas-filled microbubbles, ultrasound localization microscopy (ULM) can map the blood flow velocity with sub-diffraction spatial resolution. In this work, we present a 3D deep-brain imaging system that seamlessly integrates PACT and ULM into a single device, 3D-PAULM. Using a low ultrasound frequency of 4 MHz, 3D-PAULM is capable of imaging the brain hemodynamic functions with intact scalp and skull in a totally non-invasive manner. Using 3D-PAULM, we studied the mouse brain functions with ischemic stroke. Multi-spectral PACT, US B-mode imaging, microbubble-enhanced power Doppler (PD), and ULM were performed on the same mouse brain with intrinsic image co-registration. From the multi-modality measurements, we further quantified blood perfusion, sO2, vessel density, and flow velocity of the mouse brain, showing stroke-induced ischemia, hypoxia, and reduced blood flow. We expect that 3D-PAULM can find broad applications in studying deep brain functions on small animal models.
AB - Photoacoustic computed tomography (PACT) is a proven technology for imaging hemodynamics in deep brain of small animal models. PACT is inherently compatible with ultrasound (US) imaging, providing complementary contrast mechanisms. While PACT can quantify the brain's oxygen saturation of hemoglobin (sO2), US imaging can probe the blood flow based on the Doppler effect. Further, by tracking gas-filled microbubbles, ultrasound localization microscopy (ULM) can map the blood flow velocity with sub-diffraction spatial resolution. In this work, we present a 3D deep-brain imaging system that seamlessly integrates PACT and ULM into a single device, 3D-PAULM. Using a low ultrasound frequency of 4 MHz, 3D-PAULM is capable of imaging the brain hemodynamic functions with intact scalp and skull in a totally non-invasive manner. Using 3D-PAULM, we studied the mouse brain functions with ischemic stroke. Multi-spectral PACT, US B-mode imaging, microbubble-enhanced power Doppler (PD), and ULM were performed on the same mouse brain with intrinsic image co-registration. From the multi-modality measurements, we further quantified blood perfusion, sO2, vessel density, and flow velocity of the mouse brain, showing stroke-induced ischemia, hypoxia, and reduced blood flow. We expect that 3D-PAULM can find broad applications in studying deep brain functions on small animal models.
KW - functional imaging
KW - Photoacoustic imaging
KW - stroke
KW - ultrasound localization microscopy
UR - http://www.scopus.com/inward/record.url?scp=85207130067&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85207130067&partnerID=8YFLogxK
U2 - 10.1109/TMI.2024.3477317
DO - 10.1109/TMI.2024.3477317
M3 - Article
C2 - 39383084
AN - SCOPUS:85207130067
SN - 0278-0062
JO - IEEE transactions on medical imaging
JF - IEEE transactions on medical imaging
ER -