Multiview optical resolution photoacoustic microscopy

  • Liren Zhu
  • , Lei Li
  • , Liang Gao
  • , Lihong V. Wang

Research output: Contribution to journalArticlepeer-review

Abstract

Optical resolution photoacoustic microscopy (OR-PAM), while providing high lateral resolution, has been limited by its relatively poor acoustically determined axial resolution. Although this limitation has been tackled in recent works by using either broadband acoustic detection or nonlinear photoacoustic effects, a flexible solution with three-dimensional optical resolution in reflection mode remains desired. Herein we present a multiview OR-PAM technique. By imaging the sample from multiple view angles and reconstructing the data using a multiview deconvolution method, we have experimentally demonstrated an isotropic optical resolution in three dimensions.

Original languageEnglish (US)
Pages (from-to)217-222
Number of pages6
JournalOptica
Volume1
Issue number4
DOIs
StatePublished - 2014
Externally publishedYes

Keywords

  • Deconvolution
  • Medical biological imaging
  • Photoacoustic imaging
  • Three-dimensional microscopy

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Atomic and Molecular Physics, and Optics

Fingerprint

Dive into the research topics of 'Multiview optical resolution photoacoustic microscopy'. Together they form a unique fingerprint.

Cite this