Abstract

Sepsis, as a leading cause of death worldwide, relies on systemic inflammatory response syndrome (SIRS) criteria for its diagnosis. SIRS is highly non-specific as it relies on monitoring of patients' vitals for sepsis diagnosis, which are known to change with many confounding factors. Changes in leukocyte counts and CD64 expression levels are known specific biomarkers of pro-inflammatory host response at the onset of sepsis. Recently, we have developed a biosensor chip that can enumerate the leukocyte counts and quantify the neutrophil CD64 expression levels from a drop of blood. We were able to show improved sepsis diagnosis and prognosis in clinical studies by measuring these parameters during different times of the patients' stay in hospital. In this paper, we investigated the rate of cell capture with CD64 expression levels and used this in a multivariate computational model using artificial neural networks (ANNs) and showed improved accuracy of quantifying CD64 expression levels from the biosensor (n = 106 whole blood experiments). We found a high coefficient of determination and low error between biosensor- and flow cytometry-based neutrophil CD64 expression levels using multiple ANN training methods in comparison to those of univariate regression commonly employed. This approach can find many applications in biosensor data analytics by utilizing multiple features of the biosensor's data for output determination.

Original languageEnglish (US)
Pages (from-to)1231-1240
Number of pages10
JournalLab on a Chip
Volume18
Issue number8
DOIs
StatePublished - Jan 1 2018

Fingerprint

Biosensing Techniques
Biosensors
Sepsis
Multivariate Analysis
Systemic Inflammatory Response Syndrome
Leukocyte Count
Neutrophils
Blood
Neural networks
Flow cytometry
Physiologic Monitoring
Biomarkers
Cause of Death
Length of Stay
Flow Cytometry
Monitoring
Experiments

ASJC Scopus subject areas

  • Bioengineering
  • Biochemistry
  • Chemistry(all)
  • Biomedical Engineering

Cite this

Multivariate computational analysis of biosensor's data for improved CD64 quantification for sepsis diagnosis. / Hassan, U.; Zhu, Ruoqing; Bashir, Rashid.

In: Lab on a Chip, Vol. 18, No. 8, 01.01.2018, p. 1231-1240.

Research output: Contribution to journalArticle

@article{3fac2d31e9c248dd89f5580e82852a44,
title = "Multivariate computational analysis of biosensor's data for improved CD64 quantification for sepsis diagnosis",
abstract = "Sepsis, as a leading cause of death worldwide, relies on systemic inflammatory response syndrome (SIRS) criteria for its diagnosis. SIRS is highly non-specific as it relies on monitoring of patients' vitals for sepsis diagnosis, which are known to change with many confounding factors. Changes in leukocyte counts and CD64 expression levels are known specific biomarkers of pro-inflammatory host response at the onset of sepsis. Recently, we have developed a biosensor chip that can enumerate the leukocyte counts and quantify the neutrophil CD64 expression levels from a drop of blood. We were able to show improved sepsis diagnosis and prognosis in clinical studies by measuring these parameters during different times of the patients' stay in hospital. In this paper, we investigated the rate of cell capture with CD64 expression levels and used this in a multivariate computational model using artificial neural networks (ANNs) and showed improved accuracy of quantifying CD64 expression levels from the biosensor (n = 106 whole blood experiments). We found a high coefficient of determination and low error between biosensor- and flow cytometry-based neutrophil CD64 expression levels using multiple ANN training methods in comparison to those of univariate regression commonly employed. This approach can find many applications in biosensor data analytics by utilizing multiple features of the biosensor's data for output determination.",
author = "U. Hassan and Ruoqing Zhu and Rashid Bashir",
year = "2018",
month = "1",
day = "1",
doi = "10.1039/c8lc00108a",
language = "English (US)",
volume = "18",
pages = "1231--1240",
journal = "Lab on a Chip - Miniaturisation for Chemistry and Biology",
issn = "1473-0197",
publisher = "Royal Society of Chemistry",
number = "8",

}

TY - JOUR

T1 - Multivariate computational analysis of biosensor's data for improved CD64 quantification for sepsis diagnosis

AU - Hassan, U.

AU - Zhu, Ruoqing

AU - Bashir, Rashid

PY - 2018/1/1

Y1 - 2018/1/1

N2 - Sepsis, as a leading cause of death worldwide, relies on systemic inflammatory response syndrome (SIRS) criteria for its diagnosis. SIRS is highly non-specific as it relies on monitoring of patients' vitals for sepsis diagnosis, which are known to change with many confounding factors. Changes in leukocyte counts and CD64 expression levels are known specific biomarkers of pro-inflammatory host response at the onset of sepsis. Recently, we have developed a biosensor chip that can enumerate the leukocyte counts and quantify the neutrophil CD64 expression levels from a drop of blood. We were able to show improved sepsis diagnosis and prognosis in clinical studies by measuring these parameters during different times of the patients' stay in hospital. In this paper, we investigated the rate of cell capture with CD64 expression levels and used this in a multivariate computational model using artificial neural networks (ANNs) and showed improved accuracy of quantifying CD64 expression levels from the biosensor (n = 106 whole blood experiments). We found a high coefficient of determination and low error between biosensor- and flow cytometry-based neutrophil CD64 expression levels using multiple ANN training methods in comparison to those of univariate regression commonly employed. This approach can find many applications in biosensor data analytics by utilizing multiple features of the biosensor's data for output determination.

AB - Sepsis, as a leading cause of death worldwide, relies on systemic inflammatory response syndrome (SIRS) criteria for its diagnosis. SIRS is highly non-specific as it relies on monitoring of patients' vitals for sepsis diagnosis, which are known to change with many confounding factors. Changes in leukocyte counts and CD64 expression levels are known specific biomarkers of pro-inflammatory host response at the onset of sepsis. Recently, we have developed a biosensor chip that can enumerate the leukocyte counts and quantify the neutrophil CD64 expression levels from a drop of blood. We were able to show improved sepsis diagnosis and prognosis in clinical studies by measuring these parameters during different times of the patients' stay in hospital. In this paper, we investigated the rate of cell capture with CD64 expression levels and used this in a multivariate computational model using artificial neural networks (ANNs) and showed improved accuracy of quantifying CD64 expression levels from the biosensor (n = 106 whole blood experiments). We found a high coefficient of determination and low error between biosensor- and flow cytometry-based neutrophil CD64 expression levels using multiple ANN training methods in comparison to those of univariate regression commonly employed. This approach can find many applications in biosensor data analytics by utilizing multiple features of the biosensor's data for output determination.

UR - http://www.scopus.com/inward/record.url?scp=85045563083&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85045563083&partnerID=8YFLogxK

U2 - 10.1039/c8lc00108a

DO - 10.1039/c8lc00108a

M3 - Article

VL - 18

SP - 1231

EP - 1240

JO - Lab on a Chip - Miniaturisation for Chemistry and Biology

JF - Lab on a Chip - Miniaturisation for Chemistry and Biology

SN - 1473-0197

IS - 8

ER -