Abstract
A solvent compensation method based on flow injection analysis is used to obtain high quality nuclear magnetic resonance (NMR) spectra during solvent gradients. Using a binary solvent system containing D2O and CD3OD, NMR line broadening and chemical shift changes are observed with a 10% methanol per min solvent composition gradient. However, by creating a second equal but reverse gradient and combining the two solvent gradients before the NMR detector, the composition of solvent reaching the NMR flow cell is kept constant. We demonstrate a system using flow injection analysis of combining solvent gradients and show constant NMR spectral performance as a function of time as the combined flow has a constant solvent composition irrespective of the initial solvent gradient. Using this approach, methods can be developed to measure high quality NMR spectra during on-flow gradient LC-NMR experiments. The ultimate ability of this approach depends on the ability to compensate for the disturbance of the solvent gradient and reverse gradient by a pair of LC columns (the analytical and reverse gradient columns).
| Original language | English (US) |
|---|---|
| Pages (from-to) | 421-426 |
| Number of pages | 6 |
| Journal | Analyst |
| Volume | 128 |
| Issue number | 5 |
| DOIs | |
| State | Published - May 1 2003 |