Abstract

This letter reports characterization and calibration of silicon microcantilever heaters for operation in liquid environments. Cantilever thermomechanical noise spectra were measured in air and in water using a commercial atomic force microscope. Cantilever electrical characteristics were measured in air and water. Laser Raman thermometry measured cantilever temperature in air and water, allowing temperature calibration through comparison with electrical characteristics. Cantilever thermal conductance in water was 136-610 W/K, depending upon the cantilever type. The cantilever electrical resistance is an accurate measure of cantilever temperature in both air and water.

Original languageEnglish (US)
Article number4655544
Pages (from-to)1805-1806
Number of pages2
JournalIEEE Sensors Journal
Volume8
Issue number11
DOIs
StatePublished - Nov 2008

Keywords

  • Laser Raman thermometry
  • Liquid operation
  • Microcantilever
  • Microcantilever heater
  • Resistive thermometer

ASJC Scopus subject areas

  • Instrumentation
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Liquid operation of silicon microcantilever heaters'. Together they form a unique fingerprint.

Cite this