Abstract
A parallel detection system for electrons in the 10-800-kV energy range is described. The design is based on a 576×382 pixel charge-coupled device array operating at liquid-nitrogen temperature. This is coupled by fiber optics to a single-crystal YAG screen, which serves as the primary detection element. The performance of the system for quantitative electron diffraction and imaging studies is described. The system will record a maximum count of 16 000 per pixel with low noise at a frame rate of 2.5 frames per second.
Original language | English (US) |
---|---|
Pages (from-to) | 2102-2105 |
Number of pages | 4 |
Journal | Review of Scientific Instruments |
Volume | 59 |
Issue number | 9 |
DOIs | |
State | Published - 1988 |
Externally published | Yes |
ASJC Scopus subject areas
- Instrumentation