TY - JOUR
T1 - LABORATORY VALIDATION of A POINT-OF-CARE CARDIAC TROPONIN i ASSAY for USE in WHITE-TAILED DEER (ODOCOILEUS VIRGINIANUS)
AU - Boesch, Jordyn M.
AU - Gleed, R. D.
AU - Erb, Hollis N.
AU - Solter, Philip F.
N1 - Publisher Copyright:
© 2015 by American Association of Zoo Veterinarians.
PY - 2015/9/1
Y1 - 2015/9/1
N2 - The VetScan® i-STAT® 1 Handheld Analyzer and cardiac troponin I (cTnI) cartridges (i-STAT cTnI assay) measured greater median cTnI concentration [cTnI] in free-ranging white-tailed deer (Odocoileus virginianus) hand-injected with anesthetic drugs after physical restraint in Clover traps than in those ground-darted with the same drugs. This suggested that Clover trapping induces myocardial damage, bringing the use of this capture method under scrutiny. The purpose of this study was to confirm the validity of the i-STAT cTnI assay in deer before recommending changes in capture methods. Median [cTnI] measured by the i-STAT cTnI assay ([cTnI]i) in heparinized whole blood collected from 52 healthy, reproductively mature, female deer physically restrained in a chute was 0.01 ng/ml (10-90% percentiles: 0.00-0.03 ng/ml; minimum, maximum: 0.00, 0.07 ng/ml); [cTnI]i was 0.00 ng/ml in 42% of the deer. There was no association between [cTnI]i and either clotting or hemolytic index. [cTnI]i was 0.00 ng/ml when deer skeletal muscle homogenate was added to deer blood with [cTnI]i of 0.00 ng/ml, confirming the i-STAT cTnI assay does not detect skeletal muscle troponins. When deer cardiac muscle homogenate was serially diluted with 1) deer blood, 2) deer plasma, and 3) cow blood, [cTnI]i was directly proportional (Y intercept = -0.09, 0.7, and -0.08 ng/ml, respectively; r2 ≥ 0.97) to the fraction of homogenate in each sample. Deer cardiac muscle homogenate was diluted with deer blood to produce three samples with low, intermediate, and high [cTnI]i; serial measurements (n = 10) performed on each sample yielded coefficients of variation (CVs) of 8, 20, and 11%, respectively. Corresponding CVs when plasma was used as diluent were 13, 9, and 7%, respectively. [cTnI]i increased when plasma with a low [cTnI]i was stored at 20-24°C for 9 days. Three freeze-thaw cycles caused no systematic change in plasma [cTnI]i.
AB - The VetScan® i-STAT® 1 Handheld Analyzer and cardiac troponin I (cTnI) cartridges (i-STAT cTnI assay) measured greater median cTnI concentration [cTnI] in free-ranging white-tailed deer (Odocoileus virginianus) hand-injected with anesthetic drugs after physical restraint in Clover traps than in those ground-darted with the same drugs. This suggested that Clover trapping induces myocardial damage, bringing the use of this capture method under scrutiny. The purpose of this study was to confirm the validity of the i-STAT cTnI assay in deer before recommending changes in capture methods. Median [cTnI] measured by the i-STAT cTnI assay ([cTnI]i) in heparinized whole blood collected from 52 healthy, reproductively mature, female deer physically restrained in a chute was 0.01 ng/ml (10-90% percentiles: 0.00-0.03 ng/ml; minimum, maximum: 0.00, 0.07 ng/ml); [cTnI]i was 0.00 ng/ml in 42% of the deer. There was no association between [cTnI]i and either clotting or hemolytic index. [cTnI]i was 0.00 ng/ml when deer skeletal muscle homogenate was added to deer blood with [cTnI]i of 0.00 ng/ml, confirming the i-STAT cTnI assay does not detect skeletal muscle troponins. When deer cardiac muscle homogenate was serially diluted with 1) deer blood, 2) deer plasma, and 3) cow blood, [cTnI]i was directly proportional (Y intercept = -0.09, 0.7, and -0.08 ng/ml, respectively; r2 ≥ 0.97) to the fraction of homogenate in each sample. Deer cardiac muscle homogenate was diluted with deer blood to produce three samples with low, intermediate, and high [cTnI]i; serial measurements (n = 10) performed on each sample yielded coefficients of variation (CVs) of 8, 20, and 11%, respectively. Corresponding CVs when plasma was used as diluent were 13, 9, and 7%, respectively. [cTnI]i increased when plasma with a low [cTnI]i was stored at 20-24°C for 9 days. Three freeze-thaw cycles caused no systematic change in plasma [cTnI]i.
KW - Cardiac troponin I
KW - Odocoileus virginianus
KW - VetScan i-STAT 1 Handheld Analyzer
KW - point-of-care assay
KW - validation
KW - white-tailed deer
UR - http://www.scopus.com/inward/record.url?scp=84941208829&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84941208829&partnerID=8YFLogxK
U2 - 10.1638/2014-0131.1
DO - 10.1638/2014-0131.1
M3 - Article
C2 - 26352949
AN - SCOPUS:84941208829
SN - 1042-7260
VL - 46
SP - 468
EP - 475
JO - Journal of Zoo and Wildlife Medicine
JF - Journal of Zoo and Wildlife Medicine
IS - 3
ER -