T1 - In vitro evaluation of three intravenous fluid line warmers
AU - Lee, Rebecca A.
AU - Towle Millard, Heather A.
AU - Weil, Ann B.
AU - Lantz, Gary
AU - Constable, Peter
AU - Lescun, Timothy B.
AU - Weng, Hsin Yi
PY - 2014/6/15
Y1 - 2014/6/15
N2 - Objective-To determine in vitro output temperature differences of 3 IV fluid warmers. Design-Prospective, randomized study. Sample-3 IV fluid warmers. Procedures-Warming capabilities of a distance-dependent blood and fluid warmer marketed for human and veterinary use (product A) and a veterinary-specific distance-dependent fluid warmer (product B) were compared at 0, 4, 8, and 12 cm from the device to the test vein and at flow rates of 20, 60, 100, 140, 180, 220, 260, and 300 mL/h with room temperature (approx 22°C) fluids (phase 1). The superior warming device was compared against a distance-independent IV fluid warmer (product C) with room temperature fluids at the same flow rates (phase 2). The effect of prewarmed fluids (38°C) versus room temperature fluids was evaluated with the superior warming device from phase 2 (phase 3). Results-In phase 1, product B produced significantly warmer fluids than product A for all flow rates and distances. Both distance-dependent devices produced warmer fluid at 0 cm, compared with 4, 8, and 12 cm. In phase 2, product B produced warmer fluid than product C at 60, 100, 140, and 180 mL/h. In phase 3, there was no significant benefit to use of prewarmed fluids versus room temperature fluids. Output temperatures ≥ 36.4°C were achieved for all rates ≥ 60 mL/h. Conclusions and Clinical Relevance-Product B had superior warming capabilities. Placing the fluid warmer close to the patient is recommended. Use of prewarmed fluids had no benefit. Lower IV fluid flow rates resulted in lower output fluid temperatures.
AB - Objective-To determine in vitro output temperature differences of 3 IV fluid warmers. Design-Prospective, randomized study. Sample-3 IV fluid warmers. Procedures-Warming capabilities of a distance-dependent blood and fluid warmer marketed for human and veterinary use (product A) and a veterinary-specific distance-dependent fluid warmer (product B) were compared at 0, 4, 8, and 12 cm from the device to the test vein and at flow rates of 20, 60, 100, 140, 180, 220, 260, and 300 mL/h with room temperature (approx 22°C) fluids (phase 1). The superior warming device was compared against a distance-independent IV fluid warmer (product C) with room temperature fluids at the same flow rates (phase 2). The effect of prewarmed fluids (38°C) versus room temperature fluids was evaluated with the superior warming device from phase 2 (phase 3). Results-In phase 1, product B produced significantly warmer fluids than product A for all flow rates and distances. Both distance-dependent devices produced warmer fluid at 0 cm, compared with 4, 8, and 12 cm. In phase 2, product B produced warmer fluid than product C at 60, 100, 140, and 180 mL/h. In phase 3, there was no significant benefit to use of prewarmed fluids versus room temperature fluids. Output temperatures ≥ 36.4°C were achieved for all rates ≥ 60 mL/h. Conclusions and Clinical Relevance-Product B had superior warming capabilities. Placing the fluid warmer close to the patient is recommended. Use of prewarmed fluids had no benefit. Lower IV fluid flow rates resulted in lower output fluid temperatures.
UR - http://www.scopus.com/inward/record.url?scp=84901627085&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84901627085&partnerID=8YFLogxK
U2 - 10.2460/javma.244.12.1423
DO - 10.2460/javma.244.12.1423
M3 - Article
C2 - 24871065
AN - SCOPUS:84901627085
SN - 0003-1488
VL - 244
SP - 1423
EP - 1428
JO - Journal of the American Veterinary Medical Association
JF - Journal of the American Veterinary Medical Association
IS - 12
ER -