High-resolution microcoil 1H-NMR for mass-limited, nanoliter-volume samples

Dean L. Olson, Timothy L. Peck, Andrew G. Webb, Richard L. Magin, Jonathan V. Sweedler

Research output: Contribution to journalArticlepeer-review

Abstract

High-resolution, proton nuclear magnetic resonance (NMR) spectra of 5-nanoliter samples have been obtained with much higher mass sensitivity [signal-to-noise ratio (S/N) per micromole] than with traditional methods. Arginine and sucrose show a mean sensitivity enhancement of 130 compared to 278-microliter samples run in a 5-millimeter tube in a conventional, commercial probe. This can reduce data acquisition time by a factor of >16,000 or reduce the needed sample mass by a factor of about 130. A linewidth of 0.6 hertz was achieved on a 300-megahertz spectrometer by matching the magnetic susceptibility of the medium that surrounds the detection cell to that of the copper coil. For sucrose, the limit of detection (defined at S/N = 3) was 19 nanograms (56 picomoles) for a 1-minute data acquisition. This technique should prove useful with mass-limited samples and for use as a detector in capillary separations.