Backscatter Coefficient Estimation Using Array Transducers

Michael F. Insana, Timothy J. Hall, Larry T. Cook

Research output: Contribution to journalArticlepeer-review

Abstract

This paper describes an extension of our broadband method for estimating backscatter coefficients from random media to include data from array transducers. Four different transducer designs have now been considered: one- and two-dimensional linear arrays, annular arrays, and single-element focused pistons commonly used in mechanical sector scanners. The analysis shows that if the backscatter echo spectrum is properly normalized, the shape of the piezoelectric elements affects only the magnitude and not the frequency dependence of the backscatter coefficient estimates. Experimental data were acquired using laboratory and clinical imaging instrumentation to verify the analysis. We compared backscatter coefficients measured as a function of frequency from well-defined scattering media that were obtained using a 1-D linear array and focused piston transducers. Instrument-independent results were found that matched theoretical predictions within the measurement error between 2 and 12 MHz. We conclude from this study that accurate backscatter coefficient estimates may be easily obtained using current clinical imaging instrumentation.

Original languageEnglish (US)
Pages (from-to)714-723
Number of pages10
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Volume41
Issue number5
DOIs
StatePublished - Sep 1994
Externally publishedYes

ASJC Scopus subject areas

  • Instrumentation
  • Acoustics and Ultrasonics
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Backscatter Coefficient Estimation Using Array Transducers'. Together they form a unique fingerprint.

Cite this