Abstract

Practical clinical optical coherence tomography (OCT) systems require automatic tools for identifying and correcting flaws in OCT images. One type of flaw is the loss of image detail owing to the dispersion of the medium, which in most cases is unknown. We present an autofocus algorithm for estimating the delay line and material dispersion from OCT reflectance data, integrating a previously presented dispersion compensation algorithm to correct the data. The algorithm is based on minimizing the Renyi entropy of the corrected axial-scan image, which is a contrast-enhancement criterion. This autofocus algorithm can be used in conjunction with a high-speed, digital-signal-processor-based OCT acquisition system for rapid image correction.

Original languageEnglish (US)
Pages (from-to)3038-3046
Number of pages9
JournalApplied Optics
Volume42
Issue number16
DOIs
StatePublished - Jun 1 2003

ASJC Scopus subject areas

  • Atomic and Molecular Physics, and Optics
  • Engineering (miscellaneous)
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Autofocus algorithm for dispersion correction in optical coherence tomography'. Together they form a unique fingerprint.

Cite this