T1 - Agreement between direct and indirect blood pressure measurements obtained from anesthetized Hispaniolan Amazon parrots
AU - Acierno, Mark J.
AU - Da Cunha, Anderson
AU - Smith, Julie
AU - Tully, Thomas N.
AU - Guzman, David Sanchez Migallon
AU - Serra, Verna
AU - Mitchell, Mark A.
N1 - Copyright:
Copyright 2009 Elsevier B.V., All rights reserved.
PY - 2008/11/15
Y1 - 2008/11/15
N2 - Objective - To determine the level of agreement between direct and indirect blood pressure measurements obtained from healthy Hispaniolan Amazon parrots (Amazona ventralis) anesthetized with isoflurane. Design - Validation study. Animals - 16 healthy adult Hispaniolan Amazon parrots. Procedures - Parrots were anesthetized, and a 26-gauge, 19-mm catheter was placed percutaneously in the superficial ulnar artery for direct measurement of systolic, mean, and diastolic arterial pressures. Indirect blood pressure measurements were obtained with a Doppler ultrasonic flow detector and an oscillometric unit. The Bland-Altman method was used to compare direct and indirect blood pressure values. Results - There was substantial disagreement between direct systolic arterial blood pressure and indirect blood pressure measurements obtained with the Doppler detector from the wing (bias, 24 mm Hg; limits of agreement, -37 to 85 mm Hg) and from the leg (bias, 14 mm Hg; limits of agreement, -14 to 42 mm Hg). Attempts to obtain indirect blood pressure measurements with the oscillometric unit were unsuccessful. Conclusions and Clinical Relevance - Results suggested that there was substantial disagreement between indirect blood pressure measurements obtained with a Doppler ultrasonic flow detector in anesthetized Hispaniolan Amazon parrots and directly measured systolic arterial blood pressure.
AB - Objective - To determine the level of agreement between direct and indirect blood pressure measurements obtained from healthy Hispaniolan Amazon parrots (Amazona ventralis) anesthetized with isoflurane. Design - Validation study. Animals - 16 healthy adult Hispaniolan Amazon parrots. Procedures - Parrots were anesthetized, and a 26-gauge, 19-mm catheter was placed percutaneously in the superficial ulnar artery for direct measurement of systolic, mean, and diastolic arterial pressures. Indirect blood pressure measurements were obtained with a Doppler ultrasonic flow detector and an oscillometric unit. The Bland-Altman method was used to compare direct and indirect blood pressure values. Results - There was substantial disagreement between direct systolic arterial blood pressure and indirect blood pressure measurements obtained with the Doppler detector from the wing (bias, 24 mm Hg; limits of agreement, -37 to 85 mm Hg) and from the leg (bias, 14 mm Hg; limits of agreement, -14 to 42 mm Hg). Attempts to obtain indirect blood pressure measurements with the oscillometric unit were unsuccessful. Conclusions and Clinical Relevance - Results suggested that there was substantial disagreement between indirect blood pressure measurements obtained with a Doppler ultrasonic flow detector in anesthetized Hispaniolan Amazon parrots and directly measured systolic arterial blood pressure.
UR - http://www.scopus.com/inward/record.url?scp=57449083896&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=57449083896&partnerID=8YFLogxK
U2 - 10.2460/javma.233.10.1587
DO - 10.2460/javma.233.10.1587
M3 - Article
C2 - 19014291
AN - SCOPUS:57449083896
SN - 0003-1488
VL - 233
SP - 1587
EP - 1590
JO - Journal of the American Veterinary Medical Association
JF - Journal of the American Veterinary Medical Association
IS - 10
ER -